Welcome to Yingmei Shenghe

Engineer's Must-Have! Quick Reference Manual for Troubleshooting Vortex Flowmeters

Engineer's Must-Have! Quick Reference Manual for Troubleshooting Vortex Flowmeters


      In the daily work of field instrument engineers, vortex flowmeters are common "old friends", but they sometimes throw a "tantrum". Unstable signals, no output, large reading deviations... In the face of these unexpected situations, a clear troubleshooting approach is more efficient than blind actions. Today's quick reference manual will help you quickly identify the problem and deal with it in one move.

    Troubleshooting 1: There is flow, but the output signal is zero or too small

When encountering this situation, first check the wiring: Is the power supply normal? Are the positive and negative poles reversed? Are the signal lines broken? If the electrical connection is fine, turn your attention to the process pipeline - check if the valve is accidentally closed or if the pipeline is blocked by impurities. If all these are normal, it is likely that the on-site vibration is too strong, exceeding the vibration resistance range of the instrument.

    Troubleshooting 2: No flow, but there is still a signal output

This is a typical "zero drift". First, rule out whether there is a slight leakage in the process pipeline or if the valve is not fully closed. If the process confirms there is no flow, the problem may lie in the instrument itself: Check if the amplifier circuit board is damp or if the sensitivity adjustment is too high. At the same time, touch the pipeline to feel the vibration. Excessive mechanical vibration is the "arch-nemesis" of vortex flowmeters, which may cause false signals.

     Troubleshooting 3: The flow reading fluctuates violently

When the reading jumps, you can take three steps. First, adjust the damping time to filter the signal; second, check the flow state, whether the upstream elbows or valves are too close to the instrument, causing vortex disorder; third, confirm if the medium is in a gas-liquid two-phase state (such as liquid vaporization or gas carrying liquid). If on-site vibration cannot be avoided, consider installing a flexible pipe or replacing it with an anti-vibration type instrument. 

      Fault Four: Abnormally High or Low Accumulated Flow

This is usually caused by "dirt" or "damage". When the flow is abnormally high, check whether the electrodes or probes are entangled with impurities. When the flow is abnormally low, consider whether the inner wall of the pipe is severely scaled or the vortex generator is corroded and worn, causing a deviation in the coefficient. At this time, the sensor needs to be cleaned or the coefficient recalibrated.

Troubleshooting is like solving a case, where logic is more important than luck. Keep this quick reference manual handy. The next time your vortex flowmeter "goes on strike", I hope you will change from being "at a loss" to being "well-prepared".

  

   Follow me to learn more about the full life cycle services of flowmeters.

   Whats App: +86 13937142958

   Email: lay022985@gmail.com



Welcome to consult Yingmei

Welcome to consult, let's start to cooperate!